Calibration Procedure - KM820/V
Verifying Infra-Red Non contact thermometers
Insert the reference thermometer probe (e.g.
KM20Ref) fully into one of the holes in the Validator.
Allow the instruments to stabilise to environment
conditions.
Switch on both instruments after stabilisation
period.
Point the Infra-Red thermometer at the base of the
Validator while ensuring the nozzle is as far inside
the Validator cup as is possible. Note the variation.
(acceptable error 2.5°C)
Verifying Contact Thermometers
Insert the reference thermometer probe (e.g.
KM20Ref) fully into one of the holes in the Validator.
Insert the probe of the Instrument being veried
fully into the other hole in the Validator.
Allow the instruments to stabilise to environment
conditions.
Switch on both instruments after stabilisation
period.
Note the variation. The acceptance error will be
dependent on the specication of the unit being
veried.
Note: The time for the instruments to stabilise will be dependent on
the validator reaching its environment temperature, which at freezer
temperatures could be up to 2 hours.
Care must be exercised when handling the validator after
exposure to low temperatures.
Comark Instruments,
52 Hurricane Way,
Norwich, Norfolk NR6 6JB
United Kingdom
Telephone: +44 844 815 6599
Fax: +44 844 815 6598
Website: www.comarkinstruments.com
Stock No. 17064 Issue 6 (11/12)
© Comark Instruments
Vista de página 0
1

Resumo do Conteúdo

Página 1

Calibration Procedure - KM820/VVerifying Infra-Red Non contact thermometers• Insert the reference thermometer probe (e.g. KM20Ref) fully into on

Comentários a estes Manuais

Sem comentários